Blumenthal EZ, Serpetopoulos CN

Surv Ophthalmol. 1998 Jan-Feb;42(4):355-7.
Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel.

ABSTRACT

The focusing-rod test is considered the gold standard for determining the correct setting of the slit-lamp ocular rings. This method is, however, somewhat time-consuming, considering the time it often takes to locate a rod every time one switches between slit-lamps or slit-lamp-mounted lasers. In this article we present an alternative test, which we call the fading-slit test, that requires only a second to perform, does not involve additional equipment (such as a rod), and can be applied at any time during the slit-lamp examination. However, while the focusing-rod test determines what setting should be used, the fading slit test can verify only whether the current setting is indeed the correct one and, therefore, it is not a complete substitute for the focusing-rod test. The fading-slit test is performed as follows: while focusing on any approximately flat surface (even the center of the patient’s dome-shaped cornea would do) and using oblique illumination, gradually reduce the slit width until it disappears. Pay attention to whether the slit continuously thins, reaching a hairline width prior to its disappearance or, alternatively, whether, just before disappearance, the slit remains somewhat wide, gradually fading away, instead of thinning away. In the former case the ocular setting is correct; in the latter case it is incorrect.

המאמר המלא- כקובץ PDF

Figure from this article:

אבחנה

אתר זה נכתב ע”י פרופ’ איתן בלומנטל. באתר תמצאו אינפורמציה מעמיקה על האבחון והטיפול בגלאוקומה ובקטרקט. האינפורמציה באתר איננה תחליף לייעוץ רפואי.

פוסט קשור

PDF icon 1
Risk Factors and Incidence of Ocular Hypertension After Penetrating Keratoplasty
Faik Oruc¸oglu, MD, Eytan Z. Blumenthal, MD, Joseph Frucht-Pery, MD, and Abraham Solomon, MD Journal of Glaucoma. 2013 Feb 19. Kudret Eye Hospital, Istanbul, Turkey Department of Ophthalmology, Hadassah-Hebrew...
PDF filing cabinet
רשימת פרסומים מדעיים של פרופ' איתן בלומנטל
ORIGINAL RESEARCH PAPERS Blumenthal EZ. Adjusting the slit-lamp oculars: An unnecessary burden or a must?  Survey of Ophthalmology 1995;40(3):225-228. Abstract Full-text (PDF) Blumenthal EZ, Serpetopoulos...
PDF icon 1
Adjusting The Slit-Lamp Oculars: An Unnecessary Burden Or A Must?
Blumenthal EZ. Surv Ophthalmol. 1995 Nov-Dec;40(3):225-8. Department of Opthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjusting the slit-lamp eyepiece rings is perceived by some...
PDF icon 1
On Focusing The Slit-Lamp: Part I. An Inaccurate Ocular Setting–What Is There To Lose?
Blumenthal EZ, Serpetopoulos CN. Surv Ophthalmol. 1998 Jan-Feb;42(4):351-4. Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjustments of the diopter rings on the...
PDF icon 1
Glaucomatous-Like Cupping Associated With Slow-Growing Supra-Sellar Intracranial Lesions
Blumenthal EZ, Girkin CA, Dotan S. Neuro-Ophthalmology. 2006;30:111-115. Department of Ophthalmology, Hebrew University-Hadassah Medical Center, Jerusalem, Israel. eblumenthal@md.huji.ac.il ABSTRACT: PURPOSE:...