Blumenthal EZ, Sample PA, Berry CC, Lee AC, Girkin CA, Zangwill L, Caprioli J, Weinreb RN.

Ophthalmology. 2003 Oct;110(10):1895-902.
Glaucoma Center and Visual Function Laboratory, Dept. of Opthalmology, University of California-San Diego, 9500 Gilman Drive, La Jolla, CA 92093-0946, USA.

ABSTRACT:

PURPOSE: To quantify factors affecting test-retest variability of threshold measurements over a series of 3 serial visual fields (VF).

DESIGN: Prospective comparative observational study.

PARTICIPANTS: Forty-one normals, 10 suspects and 35 stable glaucoma patients.

METHODS: All subjects performed 3 standard and 3 short-wavelength automated perimetry (SWAP) VFs. At each VF location, severity (defined as age-corrected total deviation) and test-retest variability (TRV), defined as the standard deviation of 3 serial threshold values, were calculated. A multiple regression model (constructed separately for standard VF and SWAP) incorporated 13 factors: severity, location, eccentricity, study group, diagnosis, superior versus inferior hemifield, nasal versus temporal hemifield, one-versus-two thresholds, age, mean pupil size, pupil size variability, between-subject variation, and residual variation.

MAIN OUTCOME MEASURES: Variability in threshold sensitivity VF values.

RESULTS: Mean TRV (+/- standard deviation) for normal, suspect and glaucoma eyes, respectively, was: 1.28 +/- 0.87, 1.53 +/- 1.04 and 2.20 +/- 1.79 dB for standard VF, and 1.87 +/- 1.35, 1.86 +/- 1.24 and 2.68 +/- 1.85 dB for SWAP. The contribution of each factor to the model for standard VF and SWAP (SWAP in parentheses) were: severity 15.5% (6.9%); location 2.7% (4.1%); eccentricity 1.1% (0.64%); diagnosis 2.9% (5.9%); “superior versus inferior” hemifield 0.17% (1.7%); “nasal versus temporal” hemifield 0.06% (0.02%); one-versus-two thresholds 0.04% (0.16%); age 0.1% (0.06%); mean pupil size 0.59% (0.1%); pupil size variability 3.2% (2.8%); between-subject 8.0% (13.5%) and residual variation 61.0% (66.6%). Excluding between-subject and residual variation, the 11-factor model was able to account for less than one third of the variability seen in both standard VF and SWAP.

CONCLUSIONS: Severity of defect and between subject variation exerted the largest effect on TRV. However, even if all 11 factors could be adjusted for, it would reduce the magnitude of TRV by only 30%. More work is needed to reduce the remaining variability inherent in psychophysical testing and to better understand the intrinsic physiological variability present both in healthy and diseased eyes. It is possible that a larger number of VFs used for the calculation of TRV might further reduce the magnitude of the remaining variability found in this study.

המאמר המלא- כקובץ PDF

 

אבחנה

אתר זה נכתב ע”י פרופ’ איתן בלומנטל. באתר תמצאו אינפורמציה מעמיקה על האבחון והטיפול בגלאוקומה ובקטרקט. האינפורמציה באתר איננה תחליף לייעוץ רפואי.

פוסט קשור

PDF icon 1
Risk Factors and Incidence of Ocular Hypertension After Penetrating Keratoplasty
Faik Oruc¸oglu, MD, Eytan Z. Blumenthal, MD, Joseph Frucht-Pery, MD, and Abraham Solomon, MD Journal of Glaucoma. 2013 Feb 19. Kudret Eye Hospital, Istanbul, Turkey Department of Ophthalmology, Hadassah-Hebrew...
PDF filing cabinet
רשימת פרסומים מדעיים של פרופ' איתן בלומנטל
ORIGINAL RESEARCH PAPERS Blumenthal EZ. Adjusting the slit-lamp oculars: An unnecessary burden or a must?  Survey of Ophthalmology 1995;40(3):225-228. Abstract Full-text (PDF) Blumenthal EZ, Serpetopoulos...
PDF icon 1
Adjusting The Slit-Lamp Oculars: An Unnecessary Burden Or A Must?
Blumenthal EZ. Surv Ophthalmol. 1995 Nov-Dec;40(3):225-8. Department of Opthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjusting the slit-lamp eyepiece rings is perceived by some...
PDF icon 1
On Focusing The Slit-Lamp: Part I. An Inaccurate Ocular Setting–What Is There To Lose?
Blumenthal EZ, Serpetopoulos CN. Surv Ophthalmol. 1998 Jan-Feb;42(4):351-4. Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjustments of the diopter rings on the...
PDF icon 1
Glaucomatous-Like Cupping Associated With Slow-Growing Supra-Sellar Intracranial Lesions
Blumenthal EZ, Girkin CA, Dotan S. Neuro-Ophthalmology. 2006;30:111-115. Department of Ophthalmology, Hebrew University-Hadassah Medical Center, Jerusalem, Israel. eblumenthal@md.huji.ac.il ABSTRACT: PURPOSE:...