E.Z. Blumenthal and S. Frenkel

Invest Ophthalmol Vis Sci 2003;44: E-Abstract 3390. © 2003 ARVO 3390 – B93
Dept of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel
Commercial Relationships: E.Z. Blumenthal, Laser Diagnostic Technologies F; S. Frenkel, None.

ABSTRACT:

PURPOSE: To evaluate the inter-device reproducibility of retinal nerve fiber layer (RNFL) thickness measurements obtained with the commercially available GDx-VCC, a scanning laser polarimeter with variable (individualized) corneal compensation.

METHODS: One eye each, from three normal subjects was tested on 13 different GDx-VCC devices, on the same day, by an experienced operator. Reproducibility for each of 5 GDx parameters was calculated.

RESULTS: For each of five tested GDx parameters, the averaged standard deviation (µm), coefficient of variation and 95% confidence interval range (µm), respectively, were: TSNIT average: 3.18µm, 5.06%, 3.84µm; superior average: 3.99µm, 5.27%, 4.82µm; inferior average: 4.55µm, 6.08%, 5.50µm; TSNIT standard deviation: 1.89µm, 8.63%, 2.92µm; the Number (also referred to as nerve fiber indicator): 4.71, N/A, 5.69. For each of the 5 GDx parameters, repeated measures ANOVA was non-significant for inter-device differences.

CONCLUSIONS: With the commercially available GDx-VCC our results indicate that the RNFL measurements appear reproducible across devices.

Keywords: imaging/image analysis: clinical imaging methods (CT, FA, ICG, MRI, OCT, RTA) clinical (human) or epidemiologic studies
 

אבחנה

אתר זה נכתב ע”י פרופ’ איתן בלומנטל. באתר תמצאו אינפורמציה מעמיקה על האבחון והטיפול בגלאוקומה ובקטרקט. האינפורמציה באתר איננה תחליף לייעוץ רפואי.

פוסט קשור

PDF icon 1
Risk Factors and Incidence of Ocular Hypertension After Penetrating Keratoplasty
Faik Oruc¸oglu, MD, Eytan Z. Blumenthal, MD, Joseph Frucht-Pery, MD, and Abraham Solomon, MD Journal of Glaucoma. 2013 Feb 19. Kudret Eye Hospital, Istanbul, Turkey Department of Ophthalmology, Hadassah-Hebrew...
PDF filing cabinet
רשימת פרסומים מדעיים של פרופ' איתן בלומנטל
ORIGINAL RESEARCH PAPERS Blumenthal EZ. Adjusting the slit-lamp oculars: An unnecessary burden or a must?  Survey of Ophthalmology 1995;40(3):225-228. Abstract Full-text (PDF) Blumenthal EZ, Serpetopoulos...
PDF icon 1
Adjusting The Slit-Lamp Oculars: An Unnecessary Burden Or A Must?
Blumenthal EZ. Surv Ophthalmol. 1995 Nov-Dec;40(3):225-8. Department of Opthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjusting the slit-lamp eyepiece rings is perceived by some...
PDF icon 1
On Focusing The Slit-Lamp: Part I. An Inaccurate Ocular Setting–What Is There To Lose?
Blumenthal EZ, Serpetopoulos CN. Surv Ophthalmol. 1998 Jan-Feb;42(4):351-4. Department of Ophthalmology, Hadassah University Hospital, Jerusalem, Israel. ABSTRACT: Adjustments of the diopter rings on the...
PDF icon 1
Glaucomatous-Like Cupping Associated With Slow-Growing Supra-Sellar Intracranial Lesions
Blumenthal EZ, Girkin CA, Dotan S. Neuro-Ophthalmology. 2006;30:111-115. Department of Ophthalmology, Hebrew University-Hadassah Medical Center, Jerusalem, Israel. eblumenthal@md.huji.ac.il ABSTRACT: PURPOSE:...